Quantification of trace elements in used lubricating oil forms a vital part in monitoring engine conditions and impact on the environment. In this study, inductively coupled plasma-optical emission spectrometry (ICP-OES) was employed for the determination of Ag, Ba, Cu, Mn and Ni in used lubricating oils. Methodology was developed so as to minimize the oil's carbonaceous matter and its effect on viscosity. Accordingly, six oil sample preparation techniques (xylene dilution, detergent emulsion, microwave digestion, dry-ashing, wet-ashing and ultrasonic extraction) were investigated for their efficiency. Optimization of the factors influencing ultrasonic-assisted extraction and ICP-OES operating parameters enabled quantification of the trace metals in oils. Limits of detection (3S b /m), in the ng g-1 range, were obtained for each element of interest using each method investigated. The validity of the methodologies studied was confirmed through the analysis of quality control (QC) samples. Analyte recoveries, ranging from 48.3 to 106 %, were obtained. Evaluation of the analytical methods studied with regard to accuracy, precision, LOD, linearity, applicability for routine analysis, preparation time and cost was made. Based on these evaluations, ultrasonic extraction has a clear advantage in terms of accuracy, applicability for routine analysis, time and cost of sample preparation.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.