AbstractA status report is given of the ENEA research activities on calibration and standardisation of radon measurement techniques. A description of radon standards and experimental facilities (the ENEA Radon Chamber and the 222Rn standard generator) developed at the Casaccia Research Centre is given. The experimental procedures to test the charcoal canister radon adsorption are described. The results of investigation of the response of passive plastic detectors (CR-39) at different exposure conditions at the ENEA radon chamber are discussed: dynamic range and accuracy of passive alpha track detectors in the range from 60 to about 7000 kBq.m-3.h radon exposure. Finally a brief description is given of the experimental facility that is under development to test instruments and methods that are used to measure the emanation/exhalation of radon gas from soils and building materials.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.