This work refers to a very easy to implementate flow injection system with potentiometric detection for l-glutamate determination in food samples. The proposed procedure is based on measurement of carbon dioxide produced by decarboxylation of l-glutamate catalyzed by l-glutamate decarboxylase (E.C. 4.1.1.1.5) from Cucurbita maxima (pumpkin). The FI potentiometric system includes an enzymatic reactor with a length of 8 cm and thickness of 5 mm packed with 200 mg of a C. maxima outer layer cut in to small pieces. The proposed procedure allowed l-glutamate determinations in the concentration interval of 10-100 mmol L(-1) for an injected sample volume of 50 microL. A phosphate buffer (0.1 mol L(-1), pH 5.5) solution flowing at 1.4 mL min(-1) was used as the carrier solution in the system. The results obtained in the analysis of food samples revealed a relative error lower than 5% when compared with those provided by the spectrophometric reference procedure. The immobilized reactor retained its initial activity for 21 days. It was possible to measure 40 samples/h with the flow system proposed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.