A rapid liquid chromatography tandem-mass spectrometry method was developed for the determination of levofloxacin and its metabolite (desmethyl-levofloxacin) in human serum. Sample preparation was done using protein precipitation technique. Our method had a run time of 2.5 min and retention times of 1.6 min for all analytes. The standard curves were linear within the concentration range of 0.10 to 5.00 mg/L for levofloxacin and 0.10 to 4.99 mg/L for desmethyllevofloxacin; a correlation coefficient (R 2 ) of 0.999 and 0.998 respectively. The lower limit of quantification for both analytes was 0.10 mg/L. Within-day precision ranged from 1.4% and 2.4% for levofloxacin, 1.5% to 5% for desmethyl-levofloxacin and between-day precision ranged from 3.6% to 4.1% for levofloxacin and 0.0% to 3.3% for desmethyl-levofloxacin; whereas, accuracy ranged from 0.1% to 12.7% for levofloxacin and 0.2% to 15.6% for desmethyl-levofloxacin. This method could be a useful asset for routine therapeutic drug monitoring of levofloxacin in multi-drug resistant tuberculosis patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.