Background: Intravenous catheters are widely used but are often removed due to complications associated with catheter sleeve formation. A catheter sleeve can develop from a thrombus, and catheter-induced vascular endothelium damage may be a critical factor for thrombus formation. We investigated the effect of catheter-induced mechanical stimulation on venous endothelial cells and catheter sleeve formation and the efficacy of anti-thrombogenic technology for preventing catheter sleeve formation in vivo. Methods: We surgically implanted poly(2-methoxyethyl acrylate)-coated and uncoated catheters with and without a stylet into the right external jugular vein of a rabbit model for 14 days. Catheter sleeve formation and the ratio of residual venous endothelial cells were compared using histological examination and immunostaining with an anti-CD31 antibody, respectively. Results: Stiffening an uncoated catheter with a stylet induced catheter sleeve formation along more than two-thirds of the length of the catheter. The ratios of residual venous endothelial cells at the tip of uncoated catheters with and without a stylet were 3% and 36%, respectively. While poly(2-methoxyethyl acrylate) coating also reduced the ratio of venous endothelial cells at the tip of the stiffened catheter (12%), it prevented external thrombus and catheter sleeve formation. Conclusion: High levels of mechanical stimulation can affect catheter-related thrombosis and promote catheter sleeve formation, and anti-thrombogenic technology such as a poly(2-methoxyethyl acrylate) coating reduces thrombus formation and can prevent catheter sleeve formation on stiffened catheters. Further studies are required to determine the maximum degree of venous endothelial cell damage before catheter sleeve formation and to compare other anti-thrombogenic technologies with poly(2-methoxyethyl acrylate) for preventing catheter sleeve formation.
The high anti-thrombogenicity of 2-hydroxyethylmethacrylate (HEMA)-styrene block copolymer (HS) and argatroban-coated polyurethane tubes in a rabbit inferior vena cava model is reported. Polyurethane tubes (i.d. 3.0 mm, o.d. 4.0 mm, 1 2.5 cm) were implanted into the inferior vena cavas of 36 rabbits, weighing 2.5 to 3.5 kg, using the sleeve anastomotic technique. The animals were divided into four groups according to the type of prosthesis: uncoated tubes (Group 1, n = 9); HS-coated tubes (Group 2, n = 9); heparin-coated tubes (Group 3, n = 9); and HS/argatroban-coated tubes (Group 4, n = 9). Patency was evaluated at 1 week by ultrasonic flowmetry and microscopic examination of the grafts. Patency rates were 0 percent (0/9) in Group 1; 56 percent (5/9) in Group 2; 67 percent (6/9) in Group 3; and 100 percent (9/9) in Group 4. Groups 2 and 3 animals demonstrated statistically significantly superior patency compared to those of Group 1. Group 4 animals exhibited significantly superior anti-thrombogenicity compared to those of both Groups 1 and 2. Although Group 3 animals had an excellent patency rate, transmission and scanning electron microscopic views showed a thick protein layer and activated platelets adhering to the surface of Group 3 vessels. On the contrary, only a thin protein layer was noted in Group 4 animals. The HS-treated surface combined with the argatroban slow-release system exhibited excellent anti-thrombogenicity under venous-flow conditions.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.