A simple, precise, accurate, and rapid high-performance thin layer chromatographic method has been developed and validated for the simultaneous quantitation of flunarizine dihydrochloride and propranolol hydrochloride in a combined capsule dosage form. The method was carried out on precoated silica gel 60 F254 TLC aluminum plate, (20×10 cm2). The solvent system was ethyl acetate:methanol:glacial acetic acid in the proportion of 8:1:1, (v/v/v). Rf value for flunarizine dihydrochloride and propranolol hydrochloride was found to be 0.62±0.02 and 0.18±0.02, respectively. The linearity regression analysis for calibration showed 0.999 and 0.999 for flunarizine dihydrochloride and propranolol hydrochloride with respect to peak area and height in the concentration range of 50-350 ng/spot and 500-3500 ng/spot, respectively. Accuracy of recovery studies was found to be 98-100.28 and 99.11-99.45% for flunarizine dihydrochloride and propranolol hydrochloride, respectively. The amounts of drug in marketed formulation were 100.5 and 101.25% of flunarizine dihydrochloride and propranolol hydrochloride, respectively. The method developed can be used for routine analysis in bulk drug and capsule dosage form.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.