Recent studies have indicated that despite stringent sterilization processes, the reusable silicone laryngeal mask airway (LMA) has the potential for disease transmission through residual biological debris. As a result, a polyvinyl chloride (PVC) disposable LMA has been introduced. This randomized trial involved 138 spontaneously ventilating adult patients who underwent elective surgery requiring airway management with an LMA to determine whether there is a clinical difference between the disposable Portex® LMA (PLMA) and the standard reusable LMA in terms of ease of insertion, intra-operative cuff pressures and postoperative incidence of sore throat. There was no significant difference in first attempt insertion success rates (79% vs 84%) or difference in the incidence of postoperative sore throat observed between the two groups. Cuff pressure increases with nitrous oxide anaesthesia were significantly larger with the reusable LMA. The disposable PLMA provided a suitable airway in spontaneously ventilating patients without the risk of disease transmission inherent in a reusable device.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.