In this study, a thin coaxial antenna for microwave ablation (MWA) is proposed. A helical slot is added between two slots in order to overcome the disadvantage of the radiation range extending to the unwanted part in the conventional single-slot and double-slot applicators. By adding a helical slot, the specific absorption rate (SAR) pattern is more concentrated near the slot as compared with the conventional slot. Experiments were conducted using the simulation, and a liver phantom was made and heated up using microwaves. Based on the results on the liver phantom, experiments were conducted using swine liver which is similar to human liver. Applied frequency and microwave average input power were assumed as 2.45 GHz and 50 W (47 dBm), respectively. As a result of the experiment, return loss was −23.09 dB, and the temperature reached 60°C after 90 seconds of exposition. Based on the phantom experiments of the swine liver, necrotic lesions of the tissue at a distance of 3.5–4 cm from the microwave antenna were observed.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.