Sulfur (S) isotope ratios of thoroughly dried organic samples were measured by direct thermal decomposition in an elemental analyzer coupled to an isotope ratio mass spectrometer in continuous flow mode (EA-CF-IRMS). For organic samples of up to 13 mg weight and with total S contents of more than 10 microg, the reproducibility of the delta34S(organic) values was +/-0.4 per thousand or better. However, the delta34S values of organic samples measured directly by online EA-CF-IRMS analysis were between 0.3 and 2.9 per thousand higher than those determined on BaSO4 precipitates produced by Parr Bomb oxidation from the same sample material. Our results suggest that structural oxygen in organic samples influences the oxygen isotope ratios of the SO2 produced from organic samples. Consequently, SO2 generated from organic samples appears to have different 18O/16O ratios than SO2 generated from BaSO4 precipitates and inorganic reference materials, resulting in a deviation from the true delta34S values because of 32S16O18O contributions to mass 66. It was shown that both the amount of structural oxygen in the organic sample, and the difference of the oxygen isotope ratios between organic samples and tank O2, influenced the magnitude of the observed deviation from the true delta34S value after direct EA-CF-IRMS analysis of organic samples. Suggestions are made to correct the difference between measured delta34S(organic) and true delta34S values in order to obtain not only reproducible, but also accurate S isotope ratios for organic materials by EA-CF-IRMS.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.