This study evaluated the effect of pre-bending dynamic compression plates on fracture site compression. Recommendations of 1 to 2 mm of pre-bend have been proposed, but there does not appear to be experimental data to confirm the optimal pre-bend magnitude. Dynamic compression plating was performed on the lateral convex surface of 18 femoral analogs to fixate a simulated mid-shaft fracture. Plates with 0 mm (flat plate), 1 mm, and 2 mm of pre-bend were evaluated for their production of compression by determining the strain magnitudes for 10 equal-sized zones across the anterior cortex at the osteotomy site using digital imaging correlation. The 0 and 1 mm plates produced significantly more compression at the near cortex (p = 0.001 and p = 0.003, respectively) than the 2 mm plate. However, the 0 and 1 mm plates also created visible diastasis at the far cortex, while the 2 mm plate exhibited compression across all zones. The strain magnitudes for the 0 mm (R2 = 0.62) and 1 mm (R2 = 0.86) plates linearly and significantly decreased from the region adjacent to the plate until a region 50%–60% across the analog diameter. In contrast, the 2 mm plate exhibited uniform strains across the osteotomy site. This study demonstrates that pre-bending a dynamic compression plate 2 mm prior to fixation on a convex lateral femur provides the most compression at the far cortex. It also produces more uniform compression across the fracture when compared to 0 and 1 mm of pre-bend.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.