Objectives
The aim of this study was to assess the objective and subjective image characteristics of monoenergetic images (MEI[+]), using a noise-optimized algorithm at different kiloelectron volts (keV) compared to polyenergetic images (PEI), in patients with pancreatic ductal adenocarcinoma (PDAC).
Methods
This retrospective, institutional review board-approved study included 45 patients (18 male, 27 female; mean age 66 years; range, 42–96 years) with PDAC who had undergone a dual-energy CT (DECT) of the abdomen for staging. One standard polyenergetic image (PEI) and five MEI(+) images in 10-keV intervals, ranging from 40 to 80 keV, were reconstructed. Line-density profile analysis, as well as the contrast-to-noise ratio (CNR) of the tumor, the signal-to-noise ratio (SNR) of the regular pancreas parenchyma and the tumor, and the CNR of the three main peripancreatic vessels, was calculated. For subjective quality assessment, two readers independently assessed the images using a 5-point Likert scale. Reader reliability was evaluated using an intraclass correlation coefficient.
Results
Line-density profile analysis revealed the largest gradient in attenuation between PDAC and regular tissue in MEI(+) at 40 keV. Low-keV MEI(+)reconstructions at 40 and 50 keV increased CNR and SNR compared to PEI (40 keV: CNR 46.8 vs. 7.5; SNR
Pankreas
32.5 vs. 15.7; SNR
Lesion
13.5 vs. 8.6;
p
< 0.001). MEI(+) at 40 keV and 50 keV were consistently preferred by the observers (
p
< 0.05), showing a high intra-observer 0.937 (0.92–0.95) and inter-observer 0.911 (0.89–0.93) agreement.
Conclusion
MEI(+) reconstructions at 40 keV and 50 keV provide better objective and subjective image quality compared to conventional PEI of DECT in patients with PDAC.
Key Points
• Low-keV MEI(+) reconstructions at 40 and 50 keV increase tumor-to-pancreas contrast compared to PEI.
• Low-keV MEI(+) reconstructions improve objective and subjective image quality parameters compared to PEI.
• Dual-energy post-processing might be a valuable tool in the diagnostic workup of patients with PDAC.
Electronic supplementary material
The online version of this article (10.1007/s00330-019-06116-9) contains supplementary material, which is available to authorized users.
Background-Aortic valve calcification (AVC) is associated with cardiovascular risk factors and coronary artery calcification. We sought to determine whether AVC is associated with the presence and extent of overall plaque burden, as well as to plaque composition (calcified, mixed, and noncalcified).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.