As part of the preparation and calibration of three new National Institute of Standards and Technology (NIST) tritiated-water radioactivity Standard Reference Materials (SRMs), we have performed a comprehensive review and critical evaluation of the half-life of tritium (hydrogen-3). Twenty three experimentally-determined values of the half-life of tritium, reported between 1936 and 2000, were found. Six of these values were updated by later values. Two values were limits. Two values were deemed to be outliers. The 13 remaining values were evaluated in several ways. The results are compared with the results of other recent evaluations and all are found to be in good agreement. Our final recommended value for the half-life of tritium is the average of the adopted values from the four most recent evaluations, (4500 ± 8) d, where 8 d corresponds to one standard uncertainty.
The hypothesis that proximity to the Sun causes variation of decay constants at permille level has been tested and disproved. Repeated activity measurements of mono-radionuclide sources were performed over periods from 200 days up to four decades at 14 laboratories across the globe. Residuals from the exponential nuclear decay curves were inspected for annual oscillations. Systematic deviations from a purely exponential decay curve differ from one data set to another and are attributable to instabilities in the instrumentation and measurement conditions. The most stable activity measurements of alpha, beta-minus, electron capture, and beta-plus decaying sources set an upper limit of 0.0006% to 0.008% to the amplitude of annual oscillations in the decay rate. Oscillations in phase with Earth’s orbital distance to the Sun could not be observed within a 10−6 to 10−5 range of precision. There are also no apparent modulations over periods of weeks or months. Consequently, there is no indication of a natural impediment against sub-permille accuracy in half-life determinations, renormalisation of activity to a distant reference date, application of nuclear dating for archaeology, geo- and cosmochronology, nor in establishing the SI unit becquerel and seeking international equivalence of activity standards.
As reported at the ICRM 2011, it was discovered that the source holder used for calibrations in the NIST 4πγ ionization chamber (IC) was not stable. This has affected a large number of half-life measurement results previously reported and used in compilations of nuclear data. Corrections have been made on all of the half-life data based on the assumption that the changes to the ionization chamber response were gradual. The corrections are energy dependent and therefore radionuclide specific. This presentation will review our results and present the recommended changes in half-life values and/or uncertainties.
As part of an international measurement intercomparison of instruments used to measure atmospheric 222Rn, four participating laboratories made nearly simultaneous measurements of 222Rn activity concentration in commonly sampled, ambient air over approximately a 2 week period, and three of these four laboratories participated in the measurement comparison of 14 introduced samples with known, but undisclosed (“blind”) 222Rn activity concentration. The exercise was conducted in Bermuda in October 1991. The 222Rn activity concentrations in ambient Bermudian air over the course of the intercomparison ranged from a few hundredths of a Bq · m−3 to about 2 Bq · m−3, while the standardized sample additions covered a range from approximately 2.5 Bq · m−3 to 35 Bq · m−3. The overall uncertainty in the latter concentrations was in the general range of 10 %, approximating a 3 standard deviation uncertainty interval. The results of the intercomparison indicated that two of the laboratories were within very good agreement with the standard additions, and almost within expected statistical variations. These same two laboratories, however, at lower ambient concentrations, exhibited a systematic difference with an averaged offset of roughly 0.3 Bq · m−3. The third laboratory participating in the measurement of standardized sample additions was systematically low by about 65 % to 70 %, with respect to the standard addition which was also confirmed in their ambient air concentration measurements. The fourth laboratory, participating in only the ambient measurement part of the intercomparison, was also systematically low by at least 40 % with respect to the first two laboratories.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.