This paper presents a novel fully implantable wireless sensor system intended for long-term monitoring of hypertension patients, designed for implantation into the femoral artery with computed tomography angiography. It consists of a pressure sensor and a telemetric unit, which is wirelessly connected to an extracorporeal readout station for energy supply and data recording. The system measures intraarterial pressure at a sampling rate of 30 Hz and an accuracy of ±1.0 mmHg over a range of 30-300 mmHg, while consuming up to 300 μW. A special peel-away sheath introducer set was developed to support the implantation procedure. The system delivered stable measurements in initial animal trials in sheep, with results being in good agreement with reference sensor systems.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.