Accuracy in spinal fusion varies greatly depending on the experience of the physician. Real-time tissue feedback with diffuse reflectance spectroscopy has been shown to provide cortical breach detection using a conventional probe with two parallel fibers. In this study, Monte Carlo simulations and optical phantom experiments were conducted to investigate how angulation of the emitting fiber affects the probed volume to allow for the detection of acute breaches. Difference in intensity magnitude between cancellous and cortical spectra increased with the fiber angle, suggesting that outward angulated fibers are beneficial in acute breach scenarios. Proximity to the cortical bone could be detected best with fibers angulated at
θ
f
=
45
∘
for impending breaches between
θ
p
=
0
∘
and
θ
p
=
45
∘
. An orthopedic surgical device comprising a third fiber perpendicular to the device axis could thus cover the full impending breach range from
θ
p
=
0
∘
to
θ
p
=
90
∘
.
Background
The increased popularity of minimally invasive spinal surgery calls for a revision of guidance techniques to prevent injuries of nearby neural and vascular structures. Lipid content has previously been proposed as a distinguishing criterion for different bone tissues to provide guidance along the interface of cancellous and cortical bone. This study aims to investigate how fat is distributed throughout the spinal column to confirm or refute the suitability of lipid content for guidance purposes.
Results
Proton density fat fraction (PDFF) was assessed over all vertebral levels for six human cadavers between 53 and 92 years of age, based on fat and water MR images. According to their distance to the vertebra contour, the data points were grouped in five regions of interest (ROIs): cortical bone (−1 mm to 0 mm), pre-cortical zone (PCZ) 1–3 (0–1 mm; 1–2 mm; 2–3 mm), and cancellous bone ($$\ge $$
≥
3 mm). For PCZ1 vs. PCZ2, a significant difference in mean PDFF of between −7.59 pp and −4.39 pp on average was found. For cortical bone vs. PCZ1, a significant difference in mean PDFF of between −27.09 pp and −18.96 pp on average was found.
Conclusion
A relationship between distance from the cortical bone boundary and lipid content could be established, paving the way for guidance techniques based on fat fraction detection for spinal surgery.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.