A quantitation study using reversed phase HPLC with UV and evaporative light-scattering detector (ELSD) was conducted on 90 library standards selected from 15 small molecule combinatorial libraries (six standards from each library). This study assessed the quantitation errors using a single calibration curve for rapid purity analysis of combinatorial libraries. The average quantitation error of six standards from one library at 200 microM by UV was 13. 4%, 20.6%, and 60.3%, at 214, 220, and 254 nm, respectively. By ELSD, the average quantitation error of these six standards at 200 micro was only 7.7%. Applying this ELSD calibration curve to 84 standards from 14 structurally diverse libraries, an average quantitation error of 16.4% was obtained. The average quantitation error of all 90 standards from 15 libraries using 15 calibration curves was 18.5%.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.