A new reflectance photometer (Glucometer, Ames Division, Miles Laboratories, Elkhart, Indiana) has been developed for use with the Dextrostix reagent strip. The device measures 8 X 17 cm, weighs 330 g, uses AA batteries, and requires no warm-up. High (330 mg/dl) and low (50 mg/dl) calibrations are stored in memory and need be repeated only when: (1) a control solution (100 mg/dl) reading is in error, (2) ambient temperature changes by more than 5 degrees C, (3) a new bottle of Dextrostix is used, and (4) batteries are changed. A 60-s timer with alarm helps make timing of the procedure consistent. Nineteen insulin-dependent and one non-insulin-dependent diabetic subjects each performed a minimum of 20 blood glucose determinations over a 3-mo period in the investigators' laboratories. Simultaneous capillary blood samples were obtained for plasma glucose analysis by a comparative method. A total of 602 patient-determined glucose values ranging from 30 to 400 mg/dl were compared with simultaneous laboratory values. These values correlated significantly (r = 0.96, P less than 0.0001, y = 2.6 + 0.92x). The Glucometer-Dextrostix System is an accurate and convenient method for home blood glucose monitoring by diabetic patients.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.