IntroductionInternational resuscitation guidelines suggest to use positive end-expiratory pressure (PEEP) during manual ventilation of neonates. Aim of our study was to test the reliability of self-inflating bags (SIB) with single-use PEEP valves regarding PEEP delivery and the effect of different peak inflation pressures (PIP) and ventilation rates (VR) on the delivered PEEP.MethodsTen new single-use PEEP valves from 5 manufacturers were tested by ventilating an intubated 1kg neonatal manikin containing a lung model with a SIB that was actuated by an electromechanical plunger device. Standard settings: PIP 20cmH2O, VR 60/min, flow 8L/min. PEEP settings of 5 and 10cmH2O were studied. A second test was conducted with settings of PIP 40cmH2O and VR 40/min. The delivered PEEP was measured by a respiratory function monitor (CO2SMO+).ResultsValves from one manufacturer delivered no relevant PEEP and were excluded. The remaining valves showed a continuous decay of the delivered pressure during expiration. The median (25th and 75th percentile) delivered PEEP with standard settings was 3.4(2.7–3.8)cmH2O when set to 5cmH2O and 6.1(4.9–7.1)cmH2O when set to 10cmH2O. Increasing the PIP from 20 to 40 cmH2O led to a median (25th and 75th percentile) decrease in PEEP to 2.3(1.8–2.7)cmH2O and 4.3(3.2–4.8)cmH2O; changing VR from 60 to 40/min led to a PEEP decrease to 2.8(2.1–3.3)cmH2O and 5.0(3.5–6.2)cmH2O for both PEEP settings.ConclusionSingle-use PEEP valves do not reliably deliver the set PEEP. PIP and VR have an effect on the delivered PEEP. Operators should be aware of these limitations when manually ventilating neonates.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.