Tip geometry and placement of rotary blood pump inflow and outflow cannulae influence the dynamics of flow within the ventricle and aortic branch. Cannulation, therefore, directly influences the potential for thrombus formation and end-organ perfusion during ventricular assist device (VAD) support or cardiopulmonary bypass (CPB). The purpose of this study was to investigate the effect of various inflow/outflow cannula tip geometries and positions on ventricular and greater vessel flow patterns to evaluate ventricular washout and impact on cerebral perfusion. Transparent models of a dilated cardiomyopathic ventricle and an aortic branch were reconstructed from magnetic resonance imaging data to allow flow measurements using particle image velocimetry (PIV). The contractile function of the failing ventricle was reproduced pneumatically, and supported with a rotary pump. Flow patterns were visualized around VAD inflow cannulae, with various tip geometries placed in three positions in the ventricle. The outflow cannula was placed in the subclavian artery and at several positions in the aorta. Flow patterns were measured using PIV and used to validate an aortic flow computational fluid dynamic study. The PIV technique indicated that locating the inflow tip in the left ventricular outflow tract improved complete ventricular washout while the tip geometry had a smaller influence. However, side holes in the inflow cannula improved washout in all cases. The PIV results confirmed that the positioning and orientation of the outflow cannula in the aortic branch had a high impact on the flow pattern in the vessels, with a negative blood flow in the right carotid artery observed in some cases. Cannula placement within the ventricle had a high influence on chamber washout. The positioning of the outflow cannula directly influences the flow through the greater vessels, and may be responsible for the occasional reduction in cerebral perfusion seen in clinical CPB.
In vitro and in vivo testing showed a safe and effective function of the ReinHeart-TAH system. This has the potential to become an alternative to transplantation. However, before a first-in-man implant, chronic animal trials still have to be completed.
Approximately 100 000 cases of oxygen deficiency in the brain occur during cardiopulmonary bypass (CPB) procedures each year. In particular, perfusion of the carotid and vertebral arteries is affected. The position of the outflow cannula influences the blood flow to the cardiovascular system and thus end organ perfusion. Traditionally, the cannula returns blood into the ascending aorta. But some surgeons prefer cannulation to the right subclavian artery. A computational fluid dynamics study was initially undertaken for both approaches. The vessel model was created from real computed tomography/magnetic resonance imaging data of young healthy patients. The simulations were run with usual CPB conditions. The flow distribution for different cannula positions in the aorta was studied, as well as the impact of the cannula tip distance to vertebral artery for the subclavian position. The study presents a fast method of analyzing the flow distribution in the cardiovascular system, and can be adapted for other applications such as ventricular assist device support. It revealed that two effects cause the loss of perfusion seen clinically: a vortex under the brachiocephalic trunk and low pressure regions near the cannula jet. The results suggest that cannulation to the subclavian artery is preferred if the cannula tip is sufficiently far away from the branch of the vertebral artery. For the aortic positions, however, the cannula should be injected from the left body side.
The hybrid MCL that we developed allows variable and continuous in vitro evaluation of mechanical circulatory support devices in TAH configuration and particularly their control algorithms in response to various cardiovascular conditions. The system has been built in a modular configuration to allow testing of different types of devices and thus provides a valuable test platform prior to animal experiments.
Oxygen deficiency in the right brain is a common problem during cardiopulmonary bypass (CPB). This is linked to an insufficient perfusion of the carotid and vertebral artery. The flow to these vessels is strongly influenced by the outflow cannula position, which is traditionally located in the ascending aorta. Another approach however is to return blood via the right subclavian artery. A computational fluid dynamics (CFD) study was performed for both methods and validated by particle image velocimetry (PIV). A 3-dimensional computer aided design model of the cardiovascular (CV) system was generated from realtime computed tomography and magnetic resonance imaging data. Mesh generation (CFD) and rapid prototyping (PIV) were used for the further model creation. The simulations were performed assuming usual CPB conditions, and the same boundary conditions were applied for the PIV validation. The flow distribution was analyzed for 55 cannula positions inside the aorta and in relation to the distance between the cannula tip and the vertebral artery branch for subclavian cannulation. The study reveals that the Venturi effect due to the cannula jet appears to be the main reason for the loss in cerebral perfusion seen clinically. It provides a PIV-validated CFD method of analyzing the flow distribution in the CV system and can be transferred to other applications.
Bileaflet MHVs were demonstrated to be only slightly sensitive to rotation regarding the washout performance of the TAH. The proposed numerical washout model proved to be an adequate tool to quantitatively compare different configurations and designs of the artificial organ regarding the potential for blood stagnation where experimental measurements are limited.
Hemocompatibility of blood pumps is a crucial parameter that has to be ensured prior to in vivo testing. In contrast to rotary blood pumps, a standard for testing a pulsatile total artificial heart (TAH) has not yet been established. Therefore, a new mock circulation loop was designed to investigate hemolysis in the left ventricle of the ReinHeart TAH. Its main features are a high hemocompatibility, physiological conditions, a low priming volume, and the conduction of blood through a closed tubing system. The mock circulation loop consists of a noninvasive pressure chamber, an aortic compliance chamber, and an atrium directly connected to the ventricle. As a control pump, the clinically approved Medos-HIA ventricular assist device (VAD) was used. The pumps were operated at 120 beats per minute with an aortic pressure of 120 to 80 mm Hg and a mean atrial pressure of 10 mm Hg, generating an output flow of about 5 L/min. Heparinized porcine blood was used. A series of six identical tests were performed. A test method was established that is comparable to ASTM F 1841, which is standard practice for the assessment of hemolysis in continuous-flow blood pumps. The average normalized index of hemolysis (NIH) values of the VAD and the ReinHeart TAH were 0.018 g/100 L and 0.03 g/100 L, respectively. The standard deviation of the NIH was 0.0033 for the VAD and 0.0034 for the TAH. Furthermore, a single test with a BPX-80 Bio-Pump was performed to verify that the hemolysis induced by the mock circulation loop was negligible. The performed tests showed a good reproducibility and statistical significance. The mock circulation loop and test protocol developed in this study are valid methods to investigate the hemolysis induced by a pulsatile blood pump.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.