Real-time measurements of carbonate ion concentrations in the ocean are critical to advancing marine environmental monitoring and research into deep-sea hydrothermal activity. Herein, we report the first example of deep-sea hydrothermal field exploration using a carbonate ion-selective electrode (ISE). The novel carbonate ISE was composed of a Ni wire as substrate, carbon film as transducers and carbonate-selective membrane layers. This paper describes the preparation process of the electrode and characterises its performance via scanning electron microscopy (SEM) and electrochemical analysis. The detection limit of the electrode for CO32− is 2.821 × 10−6 mol/L, the linear response range is 1.0 × 10−5–1.0 × 10−1 mol/L and the Nernst slope was −30.4 mV/decade. In April 2021, the carbonate ISE was mounted on multi-parameter sensors with pH and Eh (redox) electrodes for the search of hydrothermal activity at the Southwest Indian Ridge. The simultaneous potential anomalies appeared at this carbonate electrode with the pH and Eh electrodes when passing through the hydrothermal field. The study of the hydrothermal field was supported by the in situ camera video and the sulphide samples. Additionally, the carbonate electrode provides enhanced information of water chemistry for the study of the hydrothermal field.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.