Purpose To assess the determinants of technical failure of magnetic resonance (MR) elastography of the liver in a large single-center study. Materials and Methods This retrospective study was approved by the institutional review board. Seven hundred eighty-one MR elastography examinations performed in 691 consecutive patients (mean age, 58 years; male patients, 434 [62.8%]) in a single center between June 2013 and August 2014 were retrospectively evaluated. MR elastography was performed at 3.0 T (n = 443) or 1.5 T (n = 338) by using a gradient-recalled-echo pulse sequence. MR elastography and anatomic image analysis were performed by two observers. Additional observers measured liver T2* and fat fraction. Technical failure was defined as no pixel value with a confidence index higher than 95% and/or no apparent shear waves imaged. Logistic regression analysis was performed to assess potential predictive factors of technical failure of MR elastography. Results The technical failure rate of MR elastography at 1.5 T was 3.5% (12 of 338), while it was higher, 15.3% (68 of 443), at 3.0 T. On the basis of univariate analysis, body mass index, liver iron deposition, massive ascites, use of 3.0 T, presence of cirrhosis, and alcoholic liver disease were all significantly associated with failure of MR elastography (P < .004); but on the basis of multivariable analysis, only body mass index, liver iron deposition, massive ascites, and use of 3.0 T were significantly associated with failure of MR elastography (P < .004). Conclusion The technical failure rate of MR elastography with a gradient-recalled-echo pulse sequence was low at 1.5 T but substantially higher at 3.0 T. Massive ascites, iron deposition, and high body mass index were additional independent factors associated with failure of MR elastography of the liver with a two-dimensional gradient-recalled-echo pulse sequence. RSNA, 2017.
Background-X-ray fluoroscopy constitutes the fundamental imaging modality for catheter visualization during interventional electrophysiology procedures. The minimal tissue discriminative capability of fluoroscopy is mitigated in part by the use of electroanatomic mapping systems and enhanced by the integration of preacquired 3-dimensional imaging of the heart with computed tomographic or magnetic resonance (MR) imaging. A more ideal paradigm might be to use intraprocedural MR imaging to directly image and guide catheter mapping procedures. Methods and Results-An MR imaging-based electroanatomic mapping system was designed to assess the feasibility of navigating catheters to the left ventricle in vivo using MR tracking of microcoils incorporated into the catheters, measuring intracardiac ventricular electrograms, and integrating this information with 3-dimensional MR angiography and myocardial delayed enhancement images to allow ventricular substrate mapping. In all animals (4 normal, and 10 chronically infarcted swine), after transseptal puncture under fluoroscopic guidance, catheters were successfully navigated to the left ventricle with MR tracking (13 to 15 frames per second) by both transseptal and retrograde aortic approaches. Electrogram artifacts related to the MR imaging gradient pulses were successfully removed with analog and digital signal processing. In all animals, it was possible to map the entire left ventricle and to project electrogram voltage amplitude maps to identify the scarred myocardium.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.