A study comparing five different cAMP detection technologies in terms of sensitivity, robustness, and feasibility for HTS is presented. In this report, the following methods are described: a nonhomogeneous DELFIA, and the homogeneous methods based on time-resolved fluorescence (HTRF), luminescent singlet oxygen channeling or ALPHAScreen, FP, and high-affinity enzyme complementation. DELFIA had the highest sensitivity, whereas ALPHAScreen and HTRF shared several advantages, including high sensitivity, broad dynamic range, and minimal reagent addition steps. For G(s)-coupled antagonist screens, we found HTRF and ALPHAScreen the more sensitive and HTS-compatible techniques.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.