The formerly reported dependence of the room air concentrations of anaesthetic volatiles and nitrous oxide from the fresh gas flow are caused by the spillage of pollutants through scavenging devices according to DIN 13260. The use of systems according to EN 740 is not only useful in new devices but must also be recommended for superseded models of anaesthetic ventilators.
Although it is not possible to calculate desflurane concentrations from the nitrous oxide concentrations, measurement of nitrous oxide as leading substance is a valid procedure to assess the exposure of the anaesthesiology workplace to nitrous oxide and volatile anaesthetics. Significant higher nitrous oxide concentrations during desflurane anaesthesia result from early extubation of patients expiring higher concentrations of nitrous oxide.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.