The overall uncertainty associated with the use of XR-RV3 films to determine skin dose in the interventional environment can realistically be estimated to be around 20% (k = 1). This uncertainty can be reduced to within 5% if carefully monitoring scanner, film, and fitting-related errors or it can easily increase to over 40% if minimal care is not taken. This work demonstrates the importance of appropriate calibration, reading, fitting, and other film-related and scan-related processes, which will help improve the accuracy of skin dose measurements in interventional procedures.
The accuracy of RDM's DACS-integrated skin dose mapping software was acceptable considering measurement uncertainties associated with Gafchromic films.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.