The aim of this study was to assess the frequencyposition function resulting from electric stimulation of electrodes in cochlear implant subjects with significant residual hearing in their nonimplanted ear. Six cochlear implant users compared the pitch of the auditory sensation produced by stimulation of an intracochlear electrode to the pitch of acoustic pure tones presented to their contralateral nonimplanted ear. Subjects were implanted with different Clarion \ electrode arrays, designed to lie close to the inner wall of the cochlea. High-resolution radiographs were used to determine the electrode positions in the cochlea. Four out of six subjects presented electrode insertions deeper than 450-. We used a two-interval (one acoustic, one electric), two-alternative forced choice protocol (2I-2AFC), asking the subject to indicate which stimulus sounded the highest in pitch. Pure tones were used as acoustic stimuli. Electric stimuli consisted of trains of biphasic pulses presented at relatively high rates [higher than 700 pulses per second (pps)]. First, all electric stimuli were balanced in loudness across electrodes. Second, acoustic pure tones, chosen to approximate roughly the pitch sensation produced by each electrode, were balanced in loudness to electric stimuli. When electrode insertion lengths were used to describe electrode positions, the pitch sensations produced by electric stimulation were found to be more than two octaves lower than predicted by Greenwood's frequency-position function. When insertion angles were used to describe electrode positions, the pitch sensations were found about one octave lower than the frequency-position function of a normal ear. The difference found between both descriptions is because of the fact that these electrode arrays were designed to lie close to the modiolus. As a consequence, the site of excitation produced at the level of the organ of Corti corresponds to a longer length than the electrode insertion length, which is used in Greenwood's function. Although exact measurements of the round window position as well as the length of the cochlea could explain the remaining one octave difference found when insertion angles were used, physiological phenomena (e.g., stimulation of the spiral ganglion cells) could also create this difference. From these data, analysis filters could be determined in sound coding strategies to match the pitch percepts elicited by electrode stimulation. This step might be of main importance for music perception and for the fitting of bilateral cochlear implants.
Using long Med-El Combi40+ electrode arrays, it is now possible to cover the whole range of the cochlea, up to about two turns. Such insertion depths have received little attention. To evaluate the contribution of deeply inserted electrodes, five Med-El cochlear implant users were tested on vowel and consonant identification tests with fittings with first one, two, and up to five apical electrodes being deactivated. In addition, subjects performed pitch-ranking experiments, using loudness-balanced stimuli, to identify electrodes creating pitch confusions. Radiographs were taken to measure each electrode insertion depth. All subjects used each modified fitting for two periods of about 3 weeks. During the experiment, the same stimulation rate and frequency range were maintained across all the fittings used for each individual subject. After each trial period the subject had to perform three consonant and three vowel identification tests. All subjects showed deep electrode insertions ranging from 605-to 720-. The two subjects with the deepest electrode insertions showed significantly increased vowel-and consonant-identification performances with fittings with the two or three most apical electrodes deactivated compared to their standard fitting with all available electrodes activated. The other three subjects did not show significant improvements in performance when one or two of their most apical electrodes were deactivated. Four out of five subjects preferred to continue use of a fitting with one or more apical electrodes deactivated. The two subjects with the deepest insertions also showed pitch confusions between their most apical electrodes. Two possible reasons for these results are discussed. One is to reduce neural interactions related to electrodes producing pitch confusions. Another is to improve the alignment of the frequency components of sounds coded by the electrical signals delivered to each electrode to the overall pitch of the auditory perception produced by the electrical stimulation of auditory nerve fibers.
The goal of this study was to evaluate, from a psychophysical standpoint, the neural spread of excitation produced by the stimulation of different types of intracochlear electrode arrays: the Ineraid, the Clarion S-Series on its own or with the Electrode Positioning System (EPS), and the Clarion HiFocus-I with the EPS. The EPS is an independent silicone part designed to bring the electrode array close to the modiolus. Forward masking was evaluated in 12 adult subjects (3 Ineraid, 4 Clarion S-Series, 3 Clarion S-Series+EPS, 3 HiFocus-I+EPS) by psychophysical experiments conducted using trains of biphasic stimuli (813 pulses per second, 307.6 micros/phase). Masker signals (+8 dB re: threshold, 300 ms) were applied to the most apical electrode. Probe signals (30 ms, 10-ms postmasker) were delivered to more basal electrodes. Masked and unmasked detection thresholds of probe signals were measured. For both Clarion HiFocus-I subjects, measurements were conducted in both monopolar and bipolar stimulus configurations. No major differences were found in forward masking between the different intracochlear electrode arrays tested in the monopolar configuration at suprathreshold levels equivalent to those used in speech-coding strategies, but significant differences were found between subjects. A significant negative correlation also was found between the level of forward masking and the consonant identification performance. These measurements showed that the neural spread of excitation was more restricted in the bipolar configuration than in the monopolar configuration for HiFocus-I subjects. It was found that CIS strategies implemented without using apical electrodes, which showed high levels of masking, could improve consonant identification.
The goal of this study was to evaluate electrical field interactions produced by the stimulation of different types of intracochlear electrodes in 12 adult subjects (three Ineraid, four Clarion S-Series, three S-Series with the electrode positioning system-EPS and two Clarion HiFocus-I with the EPS). Psychophysical measurements were conducted with biphasic stimuli (813 pulse per second, 153.8 micros/phase). "Perturbation" signals (300 ms) were applied to one electrode chosen at the middle of the array and their effects on detection thresholds of "probe" signals (30 ms) were measured on the neighbor basal electrode. Perturbation levels were set below the detection threshold of the perturbation electrode (-2 dB re threshold). Measurements were first conducted for simultaneous stimulation of the probe and of the perturbation electrodes, for monopolar for all subjects and for bipolar stimulus configurations for both Clarion HiFocus-I subjects. The tested Clarion electrodes did not present lower monopolar interactions than the Ineraid electrodes. Nevertheless, considering the shorter distance between electrodes for the Clarion than for the Ineraid, the tested Clarion electrodes might be more selective than the Ineraid. We did not find any significant monopolar electrical field-interaction differences between subjects who received the S-Series array with and without the EPS. We did not find lower interactions for both subjects who received the HiFocus-I array than for subjects who received the S-Series. Electrical field interactions were lower for bipolar than for monopolar configurations for both HiFocus-I subjects. A second set of measurements was conducted for nonsimultaneous stimulation similar to the one used in continuous interleaved sampling sound strategy. These measurements showed that interactions evaluated for simultaneous biphasic stimuli were larger than for nonsimultaneous stimulation.
Both Clarion HiFocus I and II electrode arrays were found to be placed close to the inner wall of the cochlea. Ineraid and Med-El Combi40+ electrode arrays were both placed close to the organ of Corti, the Med-El Combi40+ arrays demonstrating the deepest insertions overall. In spite of marked differences in the positions of the contacts, we did not find any correlation with speech perception performance for the different types of implants studied.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.