It may be reasonable to study whether LEC pulses from 0.25 ms to 0.30 ms in duration and up to 250 V in amplitude would increase therapeutic yield in VT termination without VF induction in humans. Contrary to the current belief, the discrepancy between defibrillation and stimulation chronaxie is not caused by different electrode size. We postulate that the time constant of the fast sodium channel reactivation may be the underlying reason.
A short (0.12-0.36 ms), high-voltage (50-100 V) field stimulus delivered from the shock coil of an implanted ICD system can safely terminate VT, especially for VT rates below 230 bpm. We believe that it would be reasonable to test the effectiveness of automatic field-stimulus therapy from implanted ICDs in VT episodes up to 230 bpm that are not susceptible to termination by antitachycardia pacing.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.