BackgroundWe aimed at assessing the factors that can influence results of the dissemination of an already validated, new generation commercial Point-of-Care Test (POCT) for detecting celiac disease (CD), in the Mediterranean area, when used in settings where it was designed to be administered, especially in countries with poor resources.MethodsPragmatic study design. Family pediatricians at their offices in Italy, nurses and pediatricians in Slovenia and Turkey at pediatricians’, schools and university primary care centers looked for CD in 3,559 (1-14 yrs), 1,480 (14-23 yrs) and 771 (1-18 yrs) asymptomatic subjects, respectively. A new generation POCT detecting IgA-tissue antitransglutaminase antibodies and IgA deficiency in a finger-tip blood drop was used. Subjects who tested positive and those suspected of having CD were referred to a Celiac Centre to undergo further investigations in order to confirm CD diagnosis. POCT Positive Predictive Value (PPV) at tertiary care (with Negative Predictive Value) and in primary care settings, and POCT and CD rates per thousand in primary care were estimated.ResultsAt tertiary care setting, PPV of the POCT and 95% CI were 89.5 (81.3-94.3) and 90 (56-98.5) with Negative Predictive Value 98.5 (94.2-99.6) and 98.7% (92-99.8) in children and adults, respectively. In primary care settings of different countries where POCT was performed by a different number of personnel, PPV ranged from 16 to 33% and the CD and POCT rates per thousand ranged from 4.77 to 1.3 and from 31.18 to 2.59, respectively.ConclusionsInterpretation of POCT results by different personnel may influence the performance of POC but dissemination of POCT is an urgent priority to be implemented among people of countries with limited resources, such as rural populations and school children.
AIMTo verify the precision and accuracy of transglutaminase antibodies (TGA) assays across Mediterranean countries.METHODSThis study involved 8 referral centres for celiac disease (CD) in 7 Mediterranean countries. A central laboratory prepared 8 kits of 7 blinded and randomized serum samples, with a titrated amount of Human TGA IgA. Each sample was analysed three times on three different days, with each centre running a total of 21 tests. The results were included in a blindly coded report form, which was sent to the coordinator centre. The coordinator estimated the mean coefficient of Variation (CoVar = σ/μ), the mean accuracy (Accur = Vobserved - Vreal) and the mean percent variation (Var% = [(Vobserved - Vreal)/Vreal] × 100).RESULTSThe analysis showed that 79.17% of the mean variation fell between -25% and +25% of the expected value, with the accuracy and precision progressively increasing with higher titres of TGA. From values 1.25 times greater than the normal cut-off, the measurements were highly reliable.CONCLUSIONTGA estimation is a crucial step for the diagnosis of CD; given its accuracy and precision, clinicians could be confident in establishing a diagnosis.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.