Background Range of motion (ROM) measurements are essential for diagnosing and evaluating upper extremity conditions. Clinical goniometry is the most commonly used methods but it is time-consuming and skill-demanding. Recent advances in human tracking algorithm suggest potential for automatic angle measuring from RGB images. It provides an attractive alternative for at-distance measuring. However, the reliability of this method has not been fully established. The purpose of this study is to evaluate if the results of algorithm are as reliable as human raters in upper limb movements. Methods Thirty healthy young adults (20 males, 10 females) participated in this study. Participants were asked to performed a 6-motion task including movement of shoulder, elbow and wrist. Images of movements were captured by commercial digital cameras. Each movement was measured by a pose tracking algorithm (OpenPose) and compared with the surgeon-measurement results. The mean differences between the two measurements were compared. Pearson correlation coefficients were used to determine the relationship. Reliability was investigated by the intra-class correlation coefficients. Results Comparing this algorithm-based method with manual measurement, the mean differences were less than 3 degrees in 5 motions (shoulder abduction: 0.51; shoulder elevation: 2.87; elbow flexion:0.38; elbow extension:0.65; wrist extension: 0.78) except wrist flexion. All the intra-class correlation coefficients were larger than 0.60. The Pearson coefficients also showed high correlations between the two measurements (p < 0.001). Conclusions Our results indicated that pose estimation is a reliable method to measure the shoulder and elbow angles, supporting RGB images for measuring joint ROM. Our results presented the possibility that patients can assess their ROM by photos taken by a digital camera. Trial registration This study was registered in the Clinical Trials Center of The First Affiliated Hospital, Sun Yat-sen University (2021–387).
The purpose of this cross-sectional study was to determine the precision and accuracy of the measurement of finger motion with a depth camera. Fifty-five healthy adult hands were included. Measurements were done with a depth camera and compared with traditional manual goniometer measurements. Repeated measuring showed that the overall repeatability and reproducibility of extension measured with the depth camera were within 3° and 4° and that of flexion were within 13° and 14°. Compared with traditional manual goniometry, biases of extension of all finger joints and flexion of metacarpophalangeal joints were less than 5°, and the average bias of flexion of proximal and distal interphalangeal joints was 29°. We conclude that the measurement of finger extension and flexion of the metacarpophalangeal joints with a depth camera was reliable, but improvement is required in the precision and accuracy of interphalangeal joint flexion.
Background: Range of motion (ROM) measurements are essential for diagnosing and evaluating upper extremity conditions. Clinical goniometry is the most commonly used methods but it is time-consuming and skill-demanding. Recent advances in human tracking algorithm suggest potential for automatic angle measuring from RGB images. It provides an attractive alternative for at-distance measuring. However, the reliability of this method has not been fully established. The purpose of this study is to evaluate if the results of algorithm are as reliable as human raters in upper limb movements.Methods: Thirty healthy young adults (20 males, 10 females) participated in this study. Participants were asked to performed a 6-motion task including movement of shoulder, elbow and wrist. Images of movements were capture by commercial digital camera. Each movement was measured by a pose tracking algorithm and compared with the surgeon-measurement results. The mean differences between the two measurements were compared. Pearson correlation coefficients were used to determine the relationship. Reliability was investigated by the intra-class correlation coefficients.Results: Comparing this algorithm-based method with manual measurement, the mean differences were less than 3 degrees in 5 motions (shoulder abduction: 0.51; shoulder elevation: 2.87; elbow flexion:0.38; elbow extension:0.65; wrist extension: 0.78) except wrist flexion. All the intra-class correlation coefficients were larger than 0.60. The Pearson coefficients also showed high correlations between the two measurements (p<0.001). Conclusions: Our results indicated that pose estimation is a reliable method to measure the shoulder and elbow angles, supporting RGB images for measuring joint ROM. Our results proved the possibility that patients can assess their ROM by photos taken by a digital camera.Trial registration: This study was registered in the Clinical Trials Center of The First Affiliated Hospital, Sun Yat-sen University (2021-387).
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.