Using a validated in vitro ureter model for laser lithotripsy, the performance of an experimental thulium fiber laser (TFL) was studied and compared to the clinical gold standard holmium:YAG laser. The holmium laser (λ = 2120 nm) was operated with standard parameters of 600 mJ, 350 μs, 6 Hz, and 270-μm-core optical fiber. The TFL (λ=1908 nm) was operated with 35 mJ, 500 μs, 150 to 500 Hz, and a 100-μm-core fiber. Urinary stones (60% calcium oxalate monohydrate/40% calcium phosphate) of uniform mass and diameter (4 to 5 mm) were laser ablated with fibers through a flexible video-ureteroscope under saline irrigation with flow rates of 22.7 and 13.7 ml/ min for the TFL and holmium laser, respectively. The temperature 3 mm from the tube's center and 1 mm above the mesh sieve was measured by a thermocouple and recorded throughout each experiment for both lasers. Total laser and operation times were recorded once all stone fragments passed through a 1.5-mm sieve. The holmium laser time measured 167±41 s (n=12). TFL times measured 111±49, 39±11, and 23±4 s, for pulse rates of 150, 300, and 500 Hz, respectively (n=12 each). Mean peak saline irrigation temperatures reached 24±1°C for holmium, and 33±3°C, 33±7°C, and 39±6°C, for TFL at pulse rates of 150, 300, and 500 Hz, respectively. To avoid thermal buildup and provide a sufficient safety margin, TFL lithotripsy should be performed with pulse rates below 500 Hz and/or increased saline irrigation rates. The TFL rapidly fragmented kidney stones due in part to its high pulse rate, high power density, high average power, and observation of reduced stone retropulsion and may provide a clinical alternative to the conventional holmium laser for lithotripsy.
Objectives The experimental Thulium fiber laser (TFL) is currently being studied as a potential alternative to the gold standard Holmium:YAG laser for lithotripsy. Recent advances in both Holmium and TFL technology allow operation at similar laser parameters for direct comparison. The use of a “dusting” mode with low pulse energy (0.2–0.4 J) and high pulse rate (50–80 Hz) settings, is gaining popularity in lithotripsy due to the desire to produce smaller residual stone fragments during ablation, capable of being spontaneously passed through the urinary tract. Methods In this study, Holmium and TFL were directly compared for ‘dusting’ using three laser groups, G1: 0.2 J/50 Hz/10 W; G2: 0.2 J/80 Hz/16 W; and G3: 0.4 J/80 Hz/32 W. Holmium laser pulse durations ranged from 200 to 350 μs, while TFL pulse durations ranged from 500 to 1,000 μs, due to technical limitations for both laser systems. An experimental setup consisting of 1 × 1 cm cuvette with 1 mm sieve was used with continuous laser operation time limited to ≤5 minutes. Calcium oxalate monohydrate stone samples with a sample size of n = 5 were used for each group, with average initial stone mass ranging from 216 to 297 mg among groups. Results Holmium laser ablation rates were lower than for TFL at all three settings (G1: 0.3 ± 0.2 vs. 0.8 ± 0.2; G2: 0.6 ± 0.1 vs. 1.0 ± 0.4; G3: 0.7 ± 0.2 vs. 1.3 ± 0.9 mg/s). The TFL also produced a greater percentage by mass of stone dust (fragments <0.5 mm) than Holmium laser. For all three settings combined, one out of 15 (7%) stones treated with Holmium laser were completely fragmented in ≤5 minutes compared to nine out of 15 (60%) stones treated with TFL. Conclusions These preliminary studies demonstrate that the TFL is a promising alternative laser for lithotripsy when operated in dusting mode, producing higher stone ablation rates and smaller stone fragments than the Holmium laser. Clinical studies are warranted. Lasers Surg. Med. 51:522–530, 2019. © 2019 Wiley Periodicals, Inc.
The Thulium fiber laser (TFL) is being explored as an alternative to the Holmium : YAG laser for lithotripsy. TFL parameters differ in several fundamental ways from Holmium laser, including smaller fiber delivery, more strongly absorbed wavelength, low pulse energy/high pulse rate operation, and more uniform temporal pulse structure. High speed imaging of laser induced bubbles was performed at 105,000 frames per second and 10 μm spatial resolution to determine influence of these laser parameters on bubble formation and needle hydrophone data was also used to measure pressure transients. The TFL was operated at 1908 nm with pulse energies of 5-65 mJ, and pulse durations of 200-1000 μs, delivered through 105-μm-core and 270-μm-core silica optical fibers. Bubble dynamics using Holmium laser at a wavelength of 2100 nm with pulse energies of 200-1000 mJ and pulse duration of 350 μs was studied, for comparison. A single, 500 μs TFL pulse produced a bubble stream extending 1200 ± 90 μm and 1070 ± 50 μm from fiber tip, with maximum bubble widths averaging 650 ± 20 μm and 870 ± 40 μm (n = 4), for 105 μm and 270 μm fibers, respectively. These observations are consistent with previous studies which reported TFL ablation stallout at working distances beyond 1.0 mm. TFL bubble dimensions were four times smaller than for Holmium laser due to lower peak power and smaller fiber diameter used. The maximum pressure transients measured 0.6 bars at 35 mJ pulse energy for TFL and 7.5 bars at 600 mJ pulse energy for Holmium laser. These fundamental studies of bubble dynamics as a function of specific laser and fiber parameters may assist with optimization of the TFL parameters for safe and efficient lithotripsy in the clinic. Image of bubble formation during fiber optic delivery of Thulium fiber laser energy in saline (35 mJ, 500 μs).
Optical fibers, consisting of 240-μm-core trunk fibers with rounded, 450-μm-diameter ball tips, are currently used during Holmium:YAG laser lithotripsy to reduce mechanical damage to the inner lining of the ureteroscope working channel during fiber insertion and prolong ureteroscope lifetime. Similarly, this study tests a smaller, 100-μm-core fiber with 300-μm-diameter ball tip during thulium fiber laser (TFL) lithotripsy. TFL was operated at a wavelength of 1908 nm, with 35-mJ pulse energy, 500-μs pulse duration, and 300-Hz pulse rate. Calcium oxalate/phosphate stone samples were weighed, laser procedure times were measured, and ablation rates were calculated for ball tip fibers, with comparison to bare tip fibers. Photographs of ball tips were taken before and after each procedure to track ball tip degradation and determine number of procedures completed before need for replacement. A high speed camera also recorded the cavitation bubble dynamics during TFL lithotripsy. Additionally, saline irrigation rates and ureteroscope deflection were measured with and without the presence of TFL fiber. There was no statistical difference (P > 0.05) between stone ablation rates for single-use ball tip fiber (1.3 AE 0.4 mg∕s) (n ¼ 10), multiple-use ball tip fiber (1.3 AE 0.5 mg∕s) (n ¼ 44), and conventional single-use bare tip fibers (1.3 AE 0.2 mg∕s) (n ¼ 10). Ball tip durability varied widely, but fibers averaged greater than four stone procedures before failure, defined by rapid decline in stone ablation rates. Mechanical damage at the front surface of the ball tip was the limiting factor in fiber lifetime. The small fiber diameter did not significantly impact ureteroscope deflection or saline flow rates. The miniature ball tip fiber may provide a cost-effective design for safe fiber insertion through the ureteroscope working channel and into the ureter without risk of instrument damage or tissue perforation, and without compromising stone ablation efficiency during TFL lithotripsy.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
10624 S. Eastern Ave., Ste. A-614
Henderson, NV 89052, USA
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.