Developing electronic medical devices is challenging. Simulations or in vivo experiments are not sufficient to obtain pertinent comparisons between potential design options. This paper presents a new artificial tool allowing objective comparisons between electronic device topologies. The main idea is to build a tool which is sensitive to targeted biological parameters only. These tools are generally called phantoms. The phantom presented in this paper is dedicated to IntraOcular Pressure (IOP) Measurement devices used in glaucoma diagnosis and treatment. It is called Biomechanical Eye Emulator (BEE). The BEE emulates the main biomechanical parameters influencing the IOP measurements. Because it is not sensitive to the living context, the BEE is the most efficient tool to investigate the best sensor design. BEE specifications are defined to be as close as possible to chosen models (humans or animals). Its efficiency is shown with a case study on rabbits. The results clearly demonstrate the BEE phantoms efficacy in providing objective assessment metrics during the sensor design process.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.