Background
Cardiac implantable electronic device-related infections (CIED) are associated with severe morbidity and mortality. Few cases have previously documented both lead endocarditis and lead rupture simultaneously.
Case summary
We describe the case of a 73-years-old man with a dual-chamber pacemaker presenting with subacute endocarditis and recurrent cholangitis. A few months prior, the patient was diagnosed with localised colon cancer and Streptococcus sanguinis lead endocarditis based on nuclear imaging. He was given prolonged antibiotic therapy and lead explantation was to be performed after sigmoidectomy. During the following weeks, his condition worsened and he was readmitted for biliary sepsis. A chest X-ray revealed, incidentally, a complete ventricular lead rupture. Pacemaker electrogram showed ventricular undersensing, loss of ventricular capture, and high impedance. As his health declined, removal of the pacemaker was deemed unreasonable and the patient died of biliary sepsis in the next few weeks.
Discussion
We describe the case of an asymptomatic intracardiac lead fracture in the setting of colon cancer and a medically managed Streptococcus lead infection. As this complication occurred during lead infection, bacterial damage may have weakened the lead over time. As illustrated by the patient’s outcomes, long-term antibiotic therapy should only be used in cases unsuitable for device removal. Complete hardware removal remains the first-line therapy in patients with CIED-related infections.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.