Failure of neonatal patient triggered ventilation may reflect a delay in delivery of flow relative to the inspiratory effort of the infant. Transmission of diaphragmatic contraction to the sensor site (patient delay) and further transmission to and within the sensing device (device delay) both contribute to the delay in triggering. Patient and device delays were studied for different sensing systems in 36 infants, 24 of whom were intubated. Device delay was long (> 40 ms) with a conventional apnoea monitor compared with sensors placed at the airway opening (2 ms), the inspiratory (12 ms) and expiratory (3 ms) pressure transducers of the ventilator, the Graseby capsule (8 ms), strain gauges (3 ms) and oesophageal pressure (6 ms). In near normal infants, the sum of patient and device delays for the latter sensors was less than 20 ms and a minor component of the total delay. However, in severe lung disease the total delay may be more than 100 ms even for airway sensors.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.