A standpipe system was developed for testing the reliability of ground water samplers. The unit consists of a stainless steel pipe 5 inches (13 centimeters) in diameter and 100 feet (30.5 meters) in height. It has 14 sampling ports from which control samples can be withdrawn at the same time and position as the samples are collected by a sampler lowered to that position. Test solutions were made in two mixing tanks, totaling 260 gallons (980 liters), by diluting the concentrate of five volatile chlorohydrocarbons in water at two levels of concentration: 10‐to‐30 and 100‐to‐200 parts per billion (micrograms per liter).
A gas chromatograph interfaced with a purge‐and‐trap system was used to perform the analyses. Comparisons of the control samples with the sampler‐collected samples have indicated that the three non‐pumping samplers had recoveries in the range of 92.4 to 103.5 percent and the three pumping samplers had recoveries ranging from 97.7 to 101.5 percent.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.