LVAD surgery in INTERMACS 1 patients is associated with remarkably good outcome considering the already very high mortality of those patients, and compared to previously reported surgical outcomes. Our study indicates that minimally invasive LVAD implantation in cardiogenic shock decreases mortality and the incidence of postoperative AEs.
The novel LVAD HM3 has already shown excellent Conformité Européenne mark trial results. Within this cohort, 1-year survival after HM3 implantation was 85%. The HM3 showed excellent midterm results with 0% stroke and 0% pump thrombosis rates 1 year after implantation.
Outflow graft obstruction in left ventricular assist devices (LVADs) is a rare complication whose clinical presentation may be subtle. We present six cases of outflow graft obstruction in patients on HeartMate 3 LVAD detected between 648 and 1,222 days on support. Detection principles are described and treatment strategies discussed. Three patients were successfully managed with stents, one underwent surgical revision, one patient died despite emergency treatment, and the last patient without symptoms was listed for heart transplantation.
Between February 2004 and December 2015, 87 exchanges of LVADs were performed at a single center. In 41 cases pump thrombosis was the reason for LVAD exchange. A total of 28 patient years (10,276 days) were analyzed. Average ICU stay was 15.8±20.4 days and average in-hospital stay 38.1±37.3 days after LVAD exchange. After thirty days the survival rate was 80.5%, 75.6% after 6 months and 70.7% one year after LVAD exchange. Out of the study cohort, three patients have successfully undergone heart transplantation. Twelve patients suffered a stroke postoperatively (29%). Twelve patients needed postoperative dialysis (29%). No technical complications of the VAD were recorded in the study group. Two patients underwent successful LVAD explantation due to myocardial recovery. One year after LVAD exchange, 14 patients underwent re-exchange due to pump thrombosis (34%). Eight patients suffered from a LVAD related infection out of which two patients were treated by pump exchange. A total of 12 patients died during the complete one year follow up of this study (29%). Four patients died in the second, two in the third and one in the fourth year after LVAD exchange. The remaining 17 patients are still ongoing on the device. It is generally feasible to treat pump thrombosis via LVAD exchange. Yet, the exchange procedure is not without risk and the risk of re-thrombosis (34%), stroke (29%), postoperative dialysis (29%) and perioperative complications remains high.
Double tunneling technique for driveline implantation leads to significantly lowered infection rates after 5 years of LVAD therapy and it is associated with a lower mortality.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.