High purity germanium (HPGe) detectors are ubiquitous in nuclear physics experiments and are also used in numerous low radioactive background detectors. The effect of the position of 60 Co and 137 Cs point sources on the shape of spectra were studied with Monte Carlo and HPGe detector measurements. We briefly confirm previous work on the position dependence of relative heights of peaks. Spectra taken with the radiation sources placed at locations around the detector were then compared using the Kolmogorov-Smirnov (K-S) goodness-of-fit test. We demonstrate that with this method the Compton continuum spectral shape has good sensitivity to the radial location of a point-source, but poor angular resolution. We conclude with a study of the position reconstruction accuracy as a function of the number of counts from the source.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.