Three LMW heparins (LMWH), one unfractionated heparin (UH), and international standards of LMWH and UH were compared in three chromogenic substrate (CS) assays and the ‘Heptest’ clotting assay. With a two-stage CS assay, linear standard curves were obtained in the 0.1–1.0 U/ml range, nearly coinciding for all preparations. With the one-stage CS assays, standard curves were curvilinear and similar for UH and the LMWH groups. In the Heptest assay, standard curves were linear for UH but not for LMWH. Mean recovery of LMWH, added to patients’ plasma samples was 70–98% for the four assays. Variation between individual recoveries was much greater with Heptest (coefficient of variation (CV) 35–44%) than with one-stage CS assays (CV 14–21 %) or two-stage CS assays (CV 7–8 %). For monitoring LMW heparin therapy, CS assays seem preferable to Heptest. The two-stage CS assay had superior accuracy, but the one-stage CS assays were easier to perform.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.