The Omega3/LHCl pixel detector readout chip comprises a matrix of 128 X 16 readout cells of 50 mu m X 500 mu m and peripheral functions with 4 distinct modes of initialization and operation, together more than 800 000 transistors. Each cell contains a complete chain of amplifier, discriminator with adjustable threshold and fast-OR output, a globally adjustable delay with local fine-tuning, coincidence logic and memory. Every cell can be individually addressed for electrical test and masking, First results have been obtained from electrical tests of a chip without detector as well as from source measurements, The electronic noise without detector is similar to 100 e(-) rms. The lowest threshold setting is close to 2000 e(-) and non-uniformity has been measured to be better than 450 e(-) rms at 5000 e(-) threshold. A timewalk of < 10 ns and a precision of < 6 ns rms on a delay of 2 mu s have been measured. The results may be improved by further optimization
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.