Purpose
This study aimed to present our experience with failures in C-TDR and revision surgery outcomes.
Materials and Methods
We retrospectively examined patients who underwent revision surgery due to the failure of C-TDR between May 2005 to March 2019. Thirteen patients (8 males and 5 females) were included in this study. The mean age was 46.1 years (range: 22–61 years), and the average follow-up period was 19.5 months (range: 12–64 months). The outcome measures of pre- and post-operative neck and arm pain using a visual analogue scale (VAS) and functional impairment were assessed using a modified Japanese Orthopedic Association (JOA) scale and the Neck Disability Index (NDI).
Results
The main complaints of patients were posterior neck pain (77%), radiculopathy (62%), and/or myelopathy (62%). The causes of failure of C-TDR were improper indications for the procedure, osteolysis and mobile implant use, inappropriate techniques, and postoperative infection. The most common surgical level was C5–6, followed by C4–5. After revision surgery, the neck and arm pain VAS (preoperative vs. postoperative: 5.46 vs. 1.31; 4.86 vs. 1.08), a modified JOA scale (14.46 vs. 16.69), and the NDI (29.77 vs. 9.31) scores were much improved.
Conclusion
C-TDR is good surgical option. However, it is very important to adhere to strict surgical indications and contraindications to avoid failure of C-TDR. The results of reoperations were good regardless of the approach. Therefore, various reoperation options could be considered in patients with failed C-TDR.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.