Purpose
To demonstrate a proton‐imaging system based on well‐established fast scintillator technology to achieve high performance with low cost and complexity, with the potential of a straightforward translation into clinical use.
Methods
The system tracks individual protons through one (X, Y) scintillating fiber tracker plane upstream and downstream of the object and into a 13‐cm ‐thick scintillating block residual energy detector. The fibers in the tracker planes are multiplexed into silicon photomultipliers (SiPMs) to reduce the number of electronics channels. The light signal from the residual energy detector is collected by 16 photomultiplier tubes (PMTs). Only four signals from the PMTs are output from each event, which allows for fast signal readout. A robust calibration method of the PMT signal to residual energy has been developed to obtain accurate proton images. The development of patient‐specific scan patterns using multiple input energies allows for an image to be produced with minimal excess dose delivered to the patient.
Results
The calibration of signals in the energy detector produces accurate residual range measurements limited by intrinsic range straggling. We measured the water‐equivalent thickness (WET) of a block of solid water (physical thickness of 6.10 mm) with a proton radiograph. The mean WET from all pixels in the block was 6.13 cm (SD 0.02 cm). The use of patient‐specific scan patterns using multiple input energies enables imaging with a compact range detector.
Conclusions
We have developed a prototype clinical proton radiography system for pretreatment imaging in proton radiation therapy. We have optimized the system for use with pencil beam scanning systems and have achieved a reduction of size and complexity compared to previous designs.
The technique of speckle visibility spectroscopy has been employed for the measurement of dynamics using coherent X-ray scattering. It is shown that the X-ray contrast within a single exposure can be related to the relaxation time of the intermediate scattering function, and this methodology is applied to the diffusion of 72 nm-radius latex spheres in glycerol. Data were collected with exposure times as short as 2 ms by employing a resonant shutter. The weak scattering present for short exposures necessitated an analysis formalism based on the spatial correlation function of individual photon charge droplets on an area detector, rather than the usual methods employed for intensity correlations. It is demonstrated that this method gives good agreement between theory and experiment and thus holds promise for extending area-detector-based coherent scattering methods to the study of faster dynamics than previously obtainable.
Verification of patient-specific proton stopping powers obtained in the patient's treatment position can be used to reduce the distal and proximal margins needed in particle beam planning. Proton radiography can be used as a pretreatment instrument to verify integrated stopping power consistency with the treatment planning CT. Although a proton radiograph is a pixel by pixel representation of integrated stopping powers, the image may also be of high enough quality and contrast to be used for patient alignment. This investigation quantifies the accuracy and image quality of a prototype proton radiography system on a clinical proton delivery system. Methods: We have developed a clinical prototype proton radiography system designed for integration into efficient clinical workflows. We tested the images obtained by this system for water-equivalent thickness (WET) accuracy, image noise, and spatial resolution. We evaluated the WET accuracy by comparing the average WET and rms error in several regions of interest (ROI) on a proton radiograph of a custom peg phantom. We measured the spatial resolution on a CATPHAN Line Pair phantom and a custom edge phantom by measuring the 10% value of the modulation transfer function (MTF). In addition, we tested the ability to detect proton range errors due to anatomical changes in a patient with a customized CIRS pediatric head phantom and inserts of varying WET placed in the posterior fossae of the brain. We took proton radiographs of the phantom with each insert in place and created difference maps between the resulting images. Integrated proton range was measured from an ROI in the difference maps. Results: We measured the WET accuracy of the proton radiographic images to be AE0.2 mm (0.33%) from known values. The spatial resolution of the images was 0.6 lp/mm on the line pair phantom and 1.13 lp/mm on the edge phantom. We were able to detect anatomical changes producing changes in WET as low as 0.6 mm.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.