An asymmetric quadrature birdcage body coil for hyperpolarized (HP) 3 He lung imaging at 1.5T is presented. The coil is designed to rest on top of the patient support and be used as a temporary insert in a clinical system. A two-part construction facilitates patient access and the asymmetric design makes maximal use of available bore space to ensure comfort. Highly homogeneous, circularly polarized RF magnetic fields are produced at 48.5 MHz using a conformal mapping method for the geometrical design, combined with an algebraic method to calculate the individual capacitance values on the birdcage coil's ladder network. Efficiency and isolation from the system's proton body coil are ensured by an integrated RF screen. The design methodology is readily applicable to other field strengths or nuclei. Improvements over existing 3 He coils were found in terms of sensitivity and transmit field homogeneity, an important feature in HP MRI. Magn Reson Med 60: 431-438, 2008.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.