We evaluated the feasibility of telerobotic microsurgical repair of corneal lacerations. The telerobotic microsurgical device consisted of a Robotic Slave Micromanipulator Unit (RSMU) coupled to a Telepresence Surgical System (TeSS). Five mm central full-thickness corneal wounds were fashioned in five enucleated rabbit eyes and repaired remotely using the telerobotic system. Five additional eyes were also repaired by hand using a standard technique. The primary outcome measure was creation of a watertight seal. All eyes in both groups maintained an intraocular pressure (IOP) of 25 mm Hg without leak. The mean repair time was 80 min (range 50-130) with telerobotic surgery compared to 8 min (range 7-9) by hand. Histological evaluation showed that suture placement was similar in robotically assisted repair and manual repair. Subjectively, the telerobotic system provided adequate three-dimensional visualization of the surgical field. The study showed that a surgeon could close standardized corneal wounds using the telerobotic system. The potential benefits of remote eye surgery include improved access, surgical teleconsultation and telementoring.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.