Observer performance tests were conducted to compare the effects of image-processing technique (unsharp mask filtering) on the diagnostic accuracy of computed radiography (CR) with storage phosphors in the detection of microcalcifications. Comparison of detectability of microcalcifications with CR and with screen-film mammography was also performed. Clusters of microcalcifications (125-250 microns in diameter) were randomly superimposed on human breast specimens. Observer performance tests were carried out with receiver operating characteristic (ROC) analysis. The area under the ROC curve and the cumulative true-positive-localization fraction were both used as indexes of performance. Observer performance experiments with nine observers indicated that the two types of screen-film images used provided higher detectability than the CR images. The detectability of microcalcifications on the CR images improved slightly with use of the unsharp masking technique. However, no statistically significant difference was found between processed and unprocessed CR images, and detectability still did not reach the level achieved with the screen-film images.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.