Inkjet printers are capable of printing at high resolution by ejecting extremely small ink drops. Established printing technology will be able to seed living cells, at micrometer resolution, in arrangements similar to biological tissues. We describe the use of a biocompatible inkjet head and our investigation of the feasibility of microseeding with living cells. Living cells are easily damaged by heat; therefore, we used an electrostatically driven inkjet system that was able to eject ink without generating significant heat. Bovine vascular endothelial cells were prepared and suspended in culture medium, and the cell suspension was used as "ink" and ejected onto culture disks. Microscopic observation showed that the endothelial cells were situated in the ejected dots in the medium, and that the number of cells in each dot was dependent on the concentration of the cell suspension and ejection frequency chosen. After the ejected cells were incubated for a few hours, they adhered to the culture disks. Using our non-heat-generating, electrostatically driven inkjet system, living cells were safely ejected onto culture disks. This microseeding technique with living cells has the potential to advance the field of tissue engineering.
The control strategy for ventricular support with a centrifugal blood pump was examined in this study. The control parameter was the pump rpm that determines pump flow. Optimum control of pump rpm that reflects the body's demand is important for long-term, effective, and safe circulatory support. Moreover, continuous, reliable monitoring of ventricular function will help successfully wean the patients from the ventricular assist device (VAD). The control strategy in this study includes determination of the target pump rpm that can provide the flow required by the body, fine-rpm-tuning to minimize deleterious effects such as suction in the ventricle, and assessment of ventricular function for successful weaning from VADs. To determine the target pump rpm, we proposed to use the relation between the native heart rate and cardiac output, and the relation between the pump rpm and centrifugal pump output. For fine-tuning of the pump rpm, the motor current waveform was used. We computed the power spectral density of the motor current waveform and calculated the ratio of the fundamental to the higher order components. When this ratio was larger than approximately 0.2, we assumed there would be a suction effect in the ventricle. As for assessment of ventricular function, we used the amplitude of the motor current waveform. The control system implemented using a DSP functioned properly in the mock circulatory loop as well as in acute animal experiments. The motor current also showed a good correlation with the ventricular pressure in acute animal experiments.
In this study, mean cell volume (MCV), mean cell hemoglobin concentration (MCHC), and mean cell hemoglobin (MCH) were measured to quantify RBC damage by rotary blood pumps. Six-hour hemolysis tests were conducted with a Bio-pump BPX-80, a Sarns 15200 roller pump, and a prototype mag-lev centrifugal pump (MedTech Heart) using fresh porcine blood circulated at 5 L/min against a 100 mm Hg head pressure. The temperature of the test and noncirculated control blood was maintained at 37 degrees C. The normalized index of hemolysis (NIH) of each pump was determined by measuring the plasma-free hemoglobin level. The MCV was measured with a Coulter counter, and MCHC was derived from total hemoglobin and hematocrit. MCH was derived from MCV and MCHC. A multivariance statistical analysis (ANOVA) revealed statistically significant differences (n = 15, P < 0.05) in MCV, MCHC, and MCH between the blood sheared by the rotary blood pumps and the nonsheared control blood. Normalized to the control blood, the Bio-pump BPX-80 showed an MCV of 1.04 +/- 0.03, an MCHC of 0.95 +/- 0.04, and an MCH of 0.98 +/- 0.02; the mag-lev MedTech Heart had an MCV of 1.02 +/- 0.02, an MCHC of 0.97 +/- 0.02, and an MCH of 0.99 +/- 0.01; and the roller pump exhibited an MCV of 1.03 +/- 0.03, an MCHC of 0.96 +/- 0.03, and an MCH of 0.99 +/- 0.01. Per 0.01 increase in NIH, the BPX-80 showed a normalized MCV change of +10.1% and a normalized MCHC change of -14.0%; the MedTech Heart demonstrated a +6.9% MCV and -9.5% MCHC change; and the roller pump had a +0.5% MCV and -0.6% MCHC change. Due to shear in the pump circuits, the RBC increased while the MCHC decreased. The likely mechanism is that older RBCs with smaller size and higher hemoglobin concentration were destroyed fast by the shear, leaving younger RBCs with larger size and lower hemoglobin concentration. Subhemolytic trauma caused the intracellular hemoglobin to decrease due to gradual hemoglobin leakage through the micropores formed in the thinned membrane. In conclusion, the rate of change in MCV and MCHC with respect to NIH change provides useful information relating to selective destruction of RBCs, while the MCH level reflects subhemolytic damage.
We have developed a tiny rotary centrifugal blood pump for the purpose of supporting circulation of children and infants. The pump is designed to provide a flow of 0.1-4.0 L/min against a head pressure of 50-120 mm Hg. The diameter of the impeller is 30 mm with six straight vanes. The impeller is supported by a hydrodynamic bearing at its center and rotated with a radial coupled magnetic driver. The bearing that supports rotation of the impeller of the tiny centrifugal blood pump is very critical to achieve durability, and clot-free and antihemolytic performance. In this study, computational fluid dynamics (CFD) analysis was performed to quantify the secondary flow through the hydrodynamic bearing at the center of the impeller and investigated the effects of bearing clearance on shear stress to optimize hemolytic performance of the pump. Two types of bearing clearance (0.1 and 0.2 mm) were studied. The wall shear stress of the 0.1-mm bearing clearance was lower than that of 0.2-mm bearing clearance at 2 L/min and 3000 rpm. This was because the axial component of the shear rate significantly decreased due to the narrower clearance even though the circumferential component of the shear rate increased. Hemolysis tests showed that the normalized index of hemolysis was reduced to 0.0076 g/100 L when the bearing clearance was reduced to 0.1 mm. It was found that the CFD prediction supported the experimental trend. The CFD is a useful tool for optimization of the hydrodynamic bearing design of the centrifugal rotary blood pump to optimize the performance of the pump in terms of mechanical effect on blood cell elements, durability of the bearing, and antithrombogenic performance.
In this study, mechanical trauma to red blood cells was evaluated by conventional hemolysis test and a newly developed cyclically reversing shear flow generator. The fresh porcine blood obtained from a local slaughterhouse was subjected to the conventional hemolysis test using a commercial centrifugal blood pump for the duration of 8 h. The measurements consisted of (i) plasma-free hemoglobin based on the standard optical measurement and (ii) the deformability of red blood cells (RBCs) using a cyclically reversing shear flow generator and microscope image acquisition system. The deformability of RBCs was expressed by the L/W value where L and W were the longer and shorter axes of the elongated RBCs' images. Although the plasma-free hemoglobin level increased with the pumping duration, the L/W remained unchanged for the duration of 8 h of pumping to indicate no alteration in the deformability. It was speculated that (i) although RBCs might have been circulated for so many times through the test pump, after each exposure to mechanical stress, RBCs might have recovered, and net effect due to shear stress-exposure time might have been small; and (ii) RBCs' deformability might be maintained near normal until sudden burst or membrane rupture, or the hemoglobin might have continuously leaked through the pores of the thinned membrane created by the mechanical stress. The deformability testing under a fluctuating shear flow could be a new method to quantify subhemolytic mechanical damage that has been accumulated in the RBCs' membrane and that may not be assessed by the conventional hemolysis test.
To elucidate the subcellular mechanism underlying the aftereffects of high-intensity dc shocks, a small pore, which mimics reversible breakdown of the cell membrane (electroporation), was incorporated into the phase-2 Luo-Rudy (L-R) model of ventricular action potentials. The pore size was set to occupy 0.15%-4.25% of the total cell membrane during the 10-ms shock. The pore was assumed to decrease after the shock exponentially with a time constant of 100-1,400 ms to simulate resealing process. In normal myocytes, the pore formation results in a delay of repolarization of the shocked action potential, which is followed by prolonged depolarization and oscillation of membrane potential like early afterdepolarization (EAD). Time- and voltage-dependent changes in the delayed rectifier K+ currents (IKr, IKs) in combination with those of L-type Ca2+ current (ICa,(L)) and ion flux through the pore (I(pore)) are responsible for the potential changes. Spontaneous excitation from the oscillation depends on activation of ICa,(L). In myocytes overloaded with Na+ and Ca2+ secondary to 90% inhibition of Na+-K+ pump, the pore formation results in a delay of repolarization of the shocked action potential, which is followed by slower cyclic depolarization in response to spontaneous release of Ca2+ from the sarcoplasmic reticulum (SR). This delayed afterdepolarization-type oscillation is abolished by complete block of Ca2+ release from the SR. These findings suggest that high-intensity electric field application will cause arrhythmogenic responses through a transient rupture of sarcolemma with different subcellular events in ventricular cells under normal and pathological conditions.
A magnetically levitated (MagLev) centrifugal blood pump (CBP) with a disposable pump head has been designed to realize a safe, easy-to-handle, reliable, and low-cost extracorporeal blood pump system. It consisted of a radial magnetic-coupled driver with a magnetic bearing having a two-degree freedom control and a disposable pump head unit with a priming volume of 24 mL. The easy on-off disposable pump head unit was made into a three-piece system consisting of the top and bottom housings, and the impeller-rotor assembly. The size and weight of the disposable pump unit were 75 mm x 45 mm and 100 g, respectively. Because the structure of the pump head unit is easily attachable and removable, the gap between the electromagnets of the stator and the target material in the rotor increased to 1.8 mm in comparison to the original integrated bearing system of 1.0 mm. The pump performance, power requirements, and controllability of the magnetic bearing revealed that from 1400 to 2400 rpm, the pump performance remained fairly unchanged. The amplitudes of the X- and Y-axis rotor oscillation increased to +/- 24 microm. The axial displacement of the rotor, 0.4 mm, toward the top housing was also observed at the pump rpm between 1400 and 2400. The axial and rotational stiffness of the bearing were 15.9 N/mm and 4.4 Nm/rad, respectively. The MagLev power was within 0.7 Watts. This study demonstrated the feasibility of a disposable, magnetically suspended CBP as the safe, reliable, easy-to-handle, low-cost extracorporeal circulation support device.
Mechanical shaft seal bearing incorporated in the centrifugal blood pumps contributes to hemolysis and thrombus formation. In addition, the problem of durability and corrosion of mechanical shaft seal bearing has been recently reported from the safety point of view. To amend the shortcomings of the blood-immersed mechanical bearings, a magnetic levitated centrifugal rotary blood pump (MedTech Dispo Model 1; Tokyo Medical and Dental University, Tokyo, Japan) has been developed for extracorporeal disposable application. In this study, the hemolytic performance of the MedTech Dispo Model 1 centrifugal blood pump system was evaluated, with special focus on the narrow blood path clearance at the magnetic bearing between rotor and stator, and on the pump housing surface roughness. A pump flow of 5 L/min against the head pressure of 100 mm Hg for 4 h was included in the hemolytic test conditions. Anticoagulated fresh porcine blood was used as a working fluid. The clearance of blood path at the magnetic bearing was in the range of 100-250 micro m. Pump housing surface roughness was controlled to be around Ra = 0.1-1.5 micro m. The lowest hemolytic results were obtained at the clearance of 250 micro m and with the polished surface (Ra = 0.1 micro m) yielding the normalized index of hemolysis (NIH) of less than 0.001 g/100 L, which was 1/5 of the Biopump BP-80 (Medtronic Inc., Minneapolis, MN, USA, and 1/4 of the BPX-80. In spite of rough surface and narrow blood path, NIH levels were less than clinically acceptable level of 0.005 g/100 L. The noncontact, levitated impeller system is useful to improve pump performance in blood environment.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.
hi@scite.ai
334 Leonard St
Brooklyn, NY 11211
Copyright © 2024 scite LLC. All rights reserved.
Made with 💙 for researchers
Part of the Research Solutions Family.