The work presented describes the modeling of a measurement bench in the presence of an implantable single-chamber pacemaker in vitro subjected to a conducted disturbance. The source of disturbance is a sinusoidal electrical signal with amplitude of 10 V; the frequency ranges studied correspond to the frequencies of the electrical distribution network (50/60 Hz) and to the frequencies of domestic applications (10/25KHz). The numerical simulations were carried out according to two methods, the first based on the finite element method using the Comsol multiphysics software, the second is based on the impedance method. The results obtained by the two methods are in good agreement, and this will allow us to calculate the voltage induced between the ends of the stimulator. These results allowed us to validate the proposed model.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.