BACKGROUND: Guidelines call for the removal of the nonvented cap (NVC) on the flushless transducer applied to the external ventricular drain (EVD) to zero the device to atmospheric pressure. Some hospitals have abandoned this practice to prevent opening the system to air. No data exist to determine the safest, most effective method of EVD zero-calibration. METHODS: A multidisciplinary team was assembled to use reflective practice to evaluate current zero-calibration of EVD practice. RESULTS: Clinical Nursing Focus showed recommendations largely out of date without detailed rationale or a high level of evidence. Manufacturer recommendations were fragmented and did not address rationale for technique. Bedside trial showed equivalence when comparing intracranial pressure (ICP) tidal, ICP after EVD zero with NVC removal, and ICP after EVD zero without NVC removal. CONCLUSION: Institutional guidelines were changed to reflect zero-calibration of EVD without NVC removal in systems that are amendable to this procedure. Further study is needed to determine best practice.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.