A novel, simple, sensitive and effective method has been developed for preconcentration of Cr (III) on functionalized Graphene oxide with covalently linked porphyrin (GO-H 2 NP) solid-phase extraction adsorbent. The stability of a chemical (GO-H 2 NP), especially in concentrated hydrochloric acid, was then used as a recycling and pre-concentration reagent. The method is based on selective chelation of Cr(III) on surfactant coated C 18 , modified with functionalized GO with covalently linked porphyrin (GO-H 2 NP). The retained ions were then eluted with 4 ml of 3 M nitric acid and determined by flame atomic absorption spectrometry (FAAS) at 283.3nm for Cr. The influence of flow rates of sample and eluent solutions, pH, breakthrough volume, effect of foreign ions on chelation and recovery were investigated. 1.5 g of surfactant coated C 18 adsorbs 40 mg of the functionalized Graphene oxide with covalently linked porphyrin (GO-H 2 NP) base which in turn can retain 20.0± 0.9 mg of each of the two ions. The limit of detection (3s) for Cr (III) was found to be 6.00 ng l -1 . The enrichment factor for both ions is 100. The mentioned method was successfully applied on the determination of Cr in different water samples. The ions were also speciated by means of a three columns system.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.