The possibility that a technical error may occur during nerve conduction studies due to ulnar nerve dislocation when the elbow is flexed has recently been suggested. We investigated normal volunteers using ultrasonography to observe the effects of ulnar nerve dislocation during elbow flexion on short-segment nerve conduction studies. We found significant conduction block in all of the subjects with ulnar nerve dislocation, and the finding was defined as a technical error caused by volume conduction. The results of the present study suggest that caution should be exercised when interpreting the results of short-segment nerve conduction studies at the across-elbow segment due to the possibility of technical error induced by ulnar nerve dislocation. GLOSSARY NCS ϭ nerve conduction study; S-NCS ϭ short-segment NCS; UNE ϭ ulnar neuropathy at the elbow. Neurology ® 2008;70:e9-e13
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.