The therapeutic advantages of scanned electron beams from medical linear accelerators are accompanied by difficulties in electron dosimetry. Owing to short-time dose inhomogeneities in the treatment field, implicit in the system, time-consuming point-by-point measurements of dose distributions are normally required. However, by triggering the measuring time synchronously with the slow scan frequency, the time necessary to record reliable dose distributions can be shortened considerably. The instrumentation for the computer-controlled dosimetry system developed for this task is described. The results of our measurements show that reproducibility in dose within +/- 1% is achievable in less than 6.3 s per point for scanned electron beams.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.