We have developed and validated a simple, rapid, and sensitive liquid chromatography analytical method employing tandem mass spectrometry (LC-MS/MS) for the determination of tolperisone, a centrally acting muscle relaxant, in human plasma. After liquid-liquid extraction with methyl t-butyl ether, chromatographic separation of tolperisone was performed using a reversed-phase Luna C(18) column (2.0mm×50mm, 5μm particles) with a mobile phase of 10mM ammonium formate buffer (pH 3.5) - methanol (12:88, v/v) and quantified by tandem mass detection in ESI positive ion mode. The flow rate of the mobile phase was 250μL/min and the retention times of tolperisone and the internal standard (IS, dibucaine) were both 0.6min. The calibration curves were linear over a range of 0.5-300ng/mL (r>0.999). The lower limit of quantification, using 200μL human plasma, was 0.5ng/mL. The mean accuracy and precision for intra- and inter-day validation of tolperisone were within acceptable limits. The LC-MS/MS method reported here showed improved sensitivity for quantification of tolperisone in human plasma compared with previously described analytical methods. Lastly, the validated method was successfully applied to a pharmacokinetic study in humans.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.