The microbial leakage at the implant-abutment interface (IAI) is one cause of peri-implant infection that puts long-term implant stability at risk. The present study compared two types of Morse taper implants in terms of sealing performance at the implanthealing abutment interface. Three implant systems, one of which exhibited a partial face-contact design (TSO) and the other two a line-contact design (TSM and BLT), were analyzed in vitro using two quantitative implant leakage measuring techniques: airinjection pressure measurement test and microbial examination. An in-house-developed device was used to determine implant leakage by recording the initial drop in pressure while injecting air through the implant. The microbial examination measured the absorbances of culture mediums in which the inoculated implants were contained. Significant difference was found between the partial face-and line-contact groups (p<0.001). Both tests revealed that partial face-contact implants are more prone to leakage than line-contact implants.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.