The non-suture techniques can be categorized into five groups based on the materials used: rings, clips, adhesives, stents and laser welding. With all these techniques a faster and less traumatic anastomosis can be made compared with sutures. However, each device is associated with technique-related complications. As a consequence, suturing continues to be the standard approach. The disadvantages of the non-suture techniques include: rigidity and a non-compliant anastomosis with rings; toxicity, leakage and aneurysm formation with adhesives; early occlusion with stents; cost, reduced strength in larger-sized vessels and demand for surgical skills with laser welding. Further refinement is needed before widespread adoption of these techniques can occur. Clips, however, may be particularly promising but long-term evaluation is required.
The use of non-penetrating clips for vascular anastomoses preserved endothelial function and structural integrity better than running sutures, although the degree of intimal hyperplasia was similar.
scite is a Brooklyn-based organization that helps researchers better discover and understand research articles through Smart Citations–citations that display the context of the citation and describe whether the article provides supporting or contrasting evidence. scite is used by students and researchers from around the world and is funded in part by the National Science Foundation and the National Institute on Drug Abuse of the National Institutes of Health.